Multimeter Command Reference
101
MEASure
The
MEASure
command subsystem configures the multimeter to perform the
specified measurement with the given range and resolution. When the multimeter is
triggered,
MEASure
makes the measurement and sends the readings to the output
buffer.
Executing
MEASure
is equivalent to configuring the multimeter with the low-level
commands shown in the following table.:
Subsystem Syntax
MEASure
:CURRent:AC? [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
:CURRent[:DC]? [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
:FREQuency? [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
:FRESistance? [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
:PERiod? [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
:RESistance? [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
:VOLTage:AC? [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
[:VOLTage[:DC]]? [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
[:VOLTage[:DC]]:RATio? [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
Command
Setting
RANGe
As specified (or AUTO).
RESolution
As specified, or as a function of range, integration
time, or NPLCs.
AC filter
([SENSe:]DET:BAND)
20 Hz - 300 kHz (medium filter)
Autozero
([SENSe:]ZERO:AUTO)
OFF if resolution setting results in NPLC <1;
ON if resolution setting results in NPLC
1
Input resistance
([SENSe:]INP:IMP:AUTO)
Applies to dc voltage and is disabled for all other
functions. 10M
for all dc voltage ranges.
Samples per trigger
(SAMP:COUN)
Trigger count
(TRIG:COUN)
Trigger delay
(TRIG:DEL)
Trigger source
(TRIG:SOUR)
1 sample
1 trigger
AUTO (Automatic delay)
IMM (trigger signal is always true)
Math function
(CALCulate:STATe)
OFF
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...