CURRent:AC:RESolution
[SENSe:]CURRent:AC:RESolution <
resolution
>
selects the resolution for AC
current measurements. See Table 3-5 on page 71.
Parameters
Comments
•
MINimum
selects the best resolution (the smallest value) for the selected range.
MAXimum selects the worst resolution (the largest value) for the selected
range.
•
You must select a range using
CURRent:AC:RANGe
before specifying
resolution. Also, only specify a resolution when making measurements on a
fixed range. Otherwise, the resolution will change to correspond with the range
selected during autoranging.
•
Specify resolution in the same units as the measurement function.
•
If autoranging is required, set the resolution using the
MIN
or
MAX
parameter.
CURRent:AC:RESolution?
[SENSe:]CURRent:AC:RESolution? [MIN|MAX]
returns one of the following
numbers to the output buffer:
•
The present current resolution selected if
MIN
or
MAX
is not specified.
•
The minimum current resolution available if
MIN
is specified.
•
The maximum current resolution available if
MAX
is specified.
Example
Query the AC Current Measurement Range
CURR:AC:RES 1E-4
Select 100
A resolution
CURR:AC:RES?
Query multimeter to return the present
resolution.
enter statement
Enter value into computer.
Parameter Name
Parameter Type
Range of Values
Default Units
<
resolution
>
numeric
resolution|
MIN|MAX
amps
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...