Multimeter Command Reference
123
CURRent[:DC]:NPLC
[SENSe:]CURRent[:DC]:NPLCycles <
number
>
sets the integration time in power
line cycles (PLCs). Values are rounded up to the nearest number of PLCs shown in
the following table.
Parameters
Comments
•
MINimum
selects 0.02 PLCs.
MAXimum
selects 100 PLCs. Setting the
integration time in power line cycles (PLCs) also sets the aperture time and the
resolution. For example, 10 PLCs sets an aperture time of 167ms (60Hz line
frequency) or 200ms (50Hz). The corresponding resolution depends on the
function and range you select.
•
The
CURR:DC:NPLC
command overrides the results of previously executed
CURRent:APERture
or
CURRent:RESolution
command.
•
The greater the number of PLCs, the greater the normal mode rejection
(and the lower the reading rate).
•
Only the 1 PLC, 10 PLC and 100 PLC settings provide normal mode rejection
of 50Hz or 60Hz power line related noise. Fractional PLC settings of 0.02 and
0.2 do not provide normal mode rejection of power line noise.
•
*RST
Condition:
10 PLC
Example
Set the DC Current Integration Time in PLCs
CURR:DC:NPLC 100
Integration time is 100 PLCs.
CURRent[:DC]:NPLC?
[SENSe:]CURRent[:DC]:NPLCycles? [MIN|MAX]
returns one of the following
numbers to the output buffer:
•
The present integration time in PLCs if
MINimum
or
MAXimum
is not
specified.
•
The minimum integration time available (0.02) if
MIN
is specified.
•
The maximum integration time available (100) if
MAX
is specified.
Example
Query the DC Current Integration Time
CURR:DC:NPLC 100
Integration time is 100 PLCs.
CURR:DC:NPLC?
Query multimeter to return integration time.
enter statement
Enter value into computer.
Parameter Name
Parameter Type
Range of Values
Default Units
<
number
>
numeric
0.02|0.2|1|10|100|MIN|MAX
PLCs
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...