Multimeter Command Reference
127
DETector:BANDwidth
[SENSe:]DETector:BANDwidth <
bw
>
selects the slow, medium or fast filter based
on the bandwidth you specify. The multimeter uses these three different filters which
enable you to either optimize low frequency accuracy or achieve faster ac settling
times on ac voltage or ac current measurements.
Parameters
-- Specifying a parameter less than 200 but greater than 20 selects the 20Hz
filter.
-- Specifying a parameter less than 20 but greater than 3 selects the 3Hz filter.
-- Specifying a parameter greater than 200 but not greater than 300E+03
(300kHz) selects the 200Hz filter.
-- Any value greater than 300kHz will cause a “Data out of range” error. The
maximum range for all three filters is 300kHz. Specify the lowest frequency
expected in the input signal. The multimeter selects the appropriate filter
based on the table below.
Comments
•
The following table lists the filter frequency range and the settling time for
making measurements.
•
The ac filter selection is stored in volatile memory and returns to the 20Hz
filter (medium) when power is removed or after a module reset. The upper
limit on all three filters is 300kHz.
•
The slow filter is 3Hz, the medium filter is 20Hz and the fast filter is 200Hz.
•
The
CONFigure
and
MEASure
commands select the 20Hz filter.
•
*RST
Condition
:
DET:BAND 20
(medium filter)
Example
Set the ac Signal Filter for Fast Measurements From 200Hz to 300kHz
DET:BAND 200
Selects the fast filter.
Parameter Name
Parameter Type
Range of Values
Default Units
<
bw
>
numeric
3|20|200|MIN|MAX
none
AC Voltage or Current
Input Frequency
AC Filter
Selected
Max Reading Rate for
Adequate Settling
3Hz to 300kHz
Slow filter
1 reading/7 seconds
20Hz to 300kHz
Medium filter
(default)
1 reading/second
200Hz to 300kHz
Fast filter
10 readings/second
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...