128
Multimeter Command Reference
DETector:BANDwidth?
[SENSe:]DETector:BANDwidth?
returns which ac filter has been selected. The
value returned is
3
,
20
or
200
. The value is sent to the output buffer.
Example
Query the Detector Bandwidth
DET:BAND 200
Select 200Hz bandwidth (fast filter).
DET:BAND?
Query multimeter to return the detector
bandwidth setting.
FREQuency:APERture
[SENSe:]FREQuency:APERture <
time
>
selects the aperture time (or gate time) for
frequency measurements.
Comments
•
Specify 0.01 (10ms) for 4½-digits, 0.1 (default, 100ms) for 5½-digits
or 1 second for 6½-digits.
•
MIN
= 0.01,
MAX
= 1.
•
*RST
Condition:
0.1 seconds
Example
Set a Frequency Aperture Time of 1 Second
FREQ:APER 1
Sets aperture time to 1 second.
FREQuency:APERture?
[SENSe:]FREQuency:APERture? [MIN|MAX]
queries the aperture time for
frequency measurements. The
MIN
parameter returns the minimum aperture
value (
0.01
); the
MAX
parameter returns the maximum aperture value (
1
).
Parameter Name
Parameter Type
Range of Values
Default Units
<
time
>
numeric
0.01|0.1|1|MIN|MAX
seconds