Multimeter Command Reference
129
FREQuency:VOLTage:RANGe
[SENSe:]FREQuency:VOLTage:RANGe <
range
>
selects the voltage range for the
signal level of frequency measurements.
Parameters
Comments
•
To select a standard measurement range, specify range as the input signal’s
maximum expected voltage. The multimeter then selects the correct range.
•
MIN
selects the minimum range available with the
VOLTage :RANGe
command: 100mV.
MAX
selects the maximum range available: 300V.
•
Specifying a fixed range disables the autorange mode set by the
FREQ:VOLT:RANG:AUTO
command.
•
The
FREQ:VOLT:RANG
command overrides the range setting from a previous
CONFigure:FREQuenc
y command.
•
*RST
Condition:
FREQ:VOLT:RANG 10
Example
Set the Voltage Range for Frequency Measurements to 100V
FREQ:VOLT:RANG 100
Voltage range for frequency measurements is
100V.
FREQuency:VOLTage:RANGe?
[SENSe:]FREQuency:VOLTage:RANGe? [MIN|MAX]
returns one of the
following numbers to the output buffer: 0.1, 1, 10, 100 or 300.
•
The present voltage range selected if
MIN
or
MAX
is not specified.
•
The minimum voltage range available (100mV) if
MIN
is specified.
•
The maximum voltage range available (300V) if
MAX
is specified.
Example
Query the Measurement Range
FREQ:VOLT:RANG 10
Select 10 V range.
FREQ:VOLT:RANG?
Query the present range.
enter statement
Enter value into computer.
Parameter Name
Parameter Type
Range of Values
Default Units
<
range
>
numeric
100mV|1V|10V|100V|300V|
MIN|MAX
volts
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...