Multimeter Command Reference
137
PERiod:VOLTage:RANGe
[SENSe:]PERiod:VOLTage:RANGe <
range
>
selects the voltage range for the
signal level of period measurements.
Parameters
Comments
•
To select a standard measurement range, specify range as the input signal’s
maximum expected voltage. The multimeter then selects the correct range.
•
MIN
selects the minimum range available with the
PERiod:VOLTage:RANGe
command: 100mV.
MAX
selects the maximum range available: 300V.
•
Specifying a fixed range disables the autorange mode set by the
PER:VOLT:RANG:AUTO
command.
•
*RST
Condition:
PER:VOLT:RANG 10
Example
Set the Voltage Range for Period Measurements to 100V
PER:VOLT:RANG 100
Voltage range is 100 V.
PERiod:VOLTage:RANGe?
[SENSe:]PERiod:VOLTage:RANGe? [MIN|MAX]
returns one of the following
numbers to the output buffer:
•
0.1, 1, 10, 100 or 300 corresponding to the range set.
•
MIN
returns 0.1.
•
MAX
returns 300.
Example
Query the Period Voltage Range
PER:VOLT:RANG?
Query the voltage range for period
measurements.
enter statement
Enter response into computer.
Parameter Name
Parameter Type
Range of Values
Default Units
<
range
>
numeric
100mV|1V|10V|100V|300V|
MIN|MAX
volts
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...