Multimeter Command Reference
147
VOLTage[:DC]:APERture
[SENSe:]VOLTage[:DC]:APERture <
number
>
sets the integration time in seconds
for dc voltage measurements. Values are rounded up to the nearest aperture time
shown in the following table.
Parameters
Comments
•
MIN
sets the aperture time to 0.333ms.
MAX
sets the aperture time
to 1.66667 seconds (60Hz) or 2 seconds (50Hz).
•
Setting the aperture time also sets the integration time in power line cycles
(PLCs) and the resolution. For example, an aperture time of 16.7ms (60Hz line
frequency) sets an integration time of 1 PLC. The corresponding resolution
depends on the function and range you select.
•
The
VOLT:APER
command overrides the results of previously executed
VOLT:NPLC
and
VOLT:RES
commands. The last command executed has
priority.
•
The greater the aperture time, the greater the normal mode rejection (and the
lower the reading rate).
•
Related Commands:
CALibration:LFRrequency
•
*RST
Condition:
VOLT:APER 0.166667
seconds (60Hz) or
VOLT:APER 0.20000
(50Hz)
Example
Set an Aperture Time of 16.7ms
VOLT:APER 16.7E-03
Aperture time is 16.7ms.
VOLTage[:DC]:APERture?
[SENSe:]VOLTage[:DC]:APERture? [MIN|MAX]
returns one of the following
numbers to the output buffer.
•
The present aperture time in seconds if
MIN
or
MAX
is not specified.
•
The minimum aperture time available (.333ms) if
MIN
is specified.
•
The maximum aperture time available (1.67s @ 60Hz; 2 s @ 50Hz) if
MAX
is
specified.
Example
Query the Aperture Time
VOLT:APER 167E-03
Aperture time is 167ms.
VOLT:APER?
Query multimeter to return aperture time.
enter statement
Enter value into computer.
Parameter Name
Parameter Type
Range of Values
Default Units
<
number
>
numeric
0.333ms|3.33ms|16.7ms|
167ms|1.67s|MIN|MAX
seconds
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...