Multimeter Command Reference
149
VOLTage[:DC]:RANGe
[SENSe:]VOLTage[:DC]:RANGe <
range
>
selects the range for DC voltage
measurements.
Parameters
Comments
•
To select a standard measurement range, specify range as the input signal’s
maximum expected voltage. The multimeter then selects the correct range.
•
MIN
selects the minimum range available with the
VOLTage:DC:RANGe
command: 100mV.
MAX
selects the maximum DC voltage range available:
300V.
•
You must select a range using
VOLTage:DC:RANGe
before specifying resolution.
•
Specifying a fixed range disables the autorange mode by setting
VOLT:DC:RANG:AUTO OFF
.
•
The
VOLT:DC:RANG
command overrides the range setting from a previous
CONFigure
command on the same function.
•
*RST
Condition:
VOLT:DC:RANG 300V
Example
Change the Range
CONF:VOLT:DC 0.85,MAX
Function: DC volts; range selected: 1V; MAX
resolution.
VOLT:DC:RANG 9
Range selected 10V; MAX resolution.
READ?
Place multimeter in wait-for-trigger state and make
measurements; send readings to output buffer.
enter statement
Enter readings into computer.
VOLTage[:DC]:RANGe?
[SENSe:]VOLTage[:DC]:RANGe? [MIN|MAX]
returns one of the following
numbers to the output buffer.
•
The present voltage range selected if
MIN
or
MAX
are not specified. Only the
ranges available with the RANGe command are returned. For example, if
CONFigure
selects the 1V range, 1.0V is the range returned.
•
The minimum voltage range available with the
VOLTage:DC:RANGe
command (100mV) if
MIN
is specified.
•
The maximum voltage range available with the
VOLTage:DC:RANGe
command (300V) if
MAX
is specified.
Example
Query the Measurement Range
VOLT:DC:RANG 1.0
Select 1V range.
VOLT:DC:RANG?
Query multimeter to return the present range.
enter statement
Enter value into computer.
Parameter Name
Parameter Type
Range of Values
Default Units
<
range
>
numeric
100mV|1V|10V|100V|300V|
MIN|MAX|DEF
volts
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...