Multimeter Command Reference
151
VOLTage[:DC]:RESolution
[SENSe:]VOLTage[:DC]:RESolution <
resolution
>
selects the resolution for DC
voltage measurements.
Parameters
Comments
•
MINimum
selects the best resolution (the smallest value) for the selected range.
MAXimum
selects the worst resolution (the largest value) for the selected range.
•
You must select a range using
VOLTage:DC:RANGe
before specifying
resolution. Also, only specify a resolution when making measurements on a
fixed range. Otherwise, the resolution will change to correspond with the range
selected during autoranging.
•
If autoranging is required, set the resolution using the
MIN
or
MAX
parameters
or select a specific integration time using
VOLTage:DC:NPLC
.
•
If necessary to achieve the specified resolution, the multimeter will increase
the integration time as needed. This command overrides the results of
previously executed
VOLTage:DC:NPLC
commands (the last command
executed has priority).
•
The
VOLT:DC:RES
command overrides the resolution setting from a previous
CONFigure:VOLT:DC:RES
command.
•
Related Commands:
CONFigure
,
VOLTage:AC:NPLC
•
*RST
Condition:
Based on the
*RST
values for the
VOLTage:NPLC
command.
Example
Change the Resolution
CONF:VOLT:DC 6.25,MAX
Function: DC volts; range selected: 10V;
MAX resolution.
VOLT:DC:RANG 0.95
Range selected: 1V; MAX resolution.
VOLT:DC:RES 3E-07
Set resolution to 0.3
V.
READ?
Place multimeter in wait-for-trigger state and make
measurements; send readings to output buffer.
enter statement
Enter readings into computer.
VOLTage[:DC]:RESolution?
[SENSe:]VOLTage[:DC]:RESolution? [MIN|MAX]
returns one of the following
numbers to the output buffer.
•
The present resolution selected if
MIN
or
MAX
is not specified. Only the
resolution values available on ranges set by the
RANGe
command are
returned.
•
The resolution with the smallest value (i.e., the best resolution) for the selected
range if
MIN
is specified.
•
The resolution with the largest value (i.e., the worst resolution) for the selected
range if
MAX
is specified.
Parameter Name
Parameter Type
Range of Values
Default Units
<
resolution
>
numeric
resolution
|MIN|MAX
volts
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...