156
Multimeter Command Reference
TRIGger
The
TRIGger
command subsystem controls the behavior of the trigger system. The
subsystem can control:
•
The number of triggers to occur before the multimeter returns to the idle state
(
TRIGger:COUNt
).
•
The delay between trigger and measurement (
TRIGger:DELay
).
•
The source of the trigger (
TRIGger:SOURce
).
Subsystem Syntax
TRIGger
:COUNt <
number
>|MIN|MAX|INFinite
:COUNt? [MIN|MAX]
:DELay <
seconds
>|MIN|MAX
:DELay? [MIN|MAX]
:DELay:AUTO OFF|ON
:DELay:AUTO?
:SOURce BUS|IMMediate|EXTernal|TTLTrg0-7
:SOURce?
:COUNt
TRIGger:COUNt <
number
>
sets the number of triggers to be issued.
Parameters
Comments
•
MIN
selects 1 trigger.
MAX
selects 50,000 triggers.
•
If
MAX
or
50,000
is specified for the number parameter, the command will be
accepted. If you initiate measurements with an
INITiate
command, an
“Insufficient memory” error occurs to show that this generates too many
readings to store in memory. However, you can use the
READ?
command to
return the readings to the output buffer and retrieve them with your controller.
The
READ?
command is a combined
INITiate
and
FETCh?
command.
•
CONFigure
and
MEASure
set the trigger count to 1.
•
*RST
Condition:
TRIG:COUN 1
Example
Set the Trigger Count
CONF:VOLT:DC
Function: DC voltage.
TRIG:SOUR EXT
Trigger source is “Trig” BNC on multimeter
front panel.
TRIG:COUN 10
Multimeter will accept 10 external triggers
(one measurement is taken with each trigger).
READ?
Place multimeter in wait-for-trigger state;
make measurement when external trigger is
received; send readings to output buffer.
enter statement
Enter readings into computer.
Parameter Name
Parameter Type
Range of Values
Default Units
<
number
>
numeric
1 through 50,000|
MIN|MAX|INF
none
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...