Multimeter Command Reference
159
:DELay:AUTO?
TRIGger:DELay:AUTO?
returns a number to show whether the automatic trigger
delay mode is on or off: “
1
” =
ON
, “
0
” =
OFF
. The number is sent to the output
buffer.
Example
Query the Trigger Delay Mode
TRIG:DEL:AUTO OFF
Disable automatic trigger delay.
TRIG:DEL:AUTO?
Query multimeter to return trigger delay mode.
enter statement
Enter value into computer.
Default Trigger Delays for DC Voltage and DC Current (all ranges):
Integration Time
Trigger Delay
NPLC
1
1.5ms
NPLC <1
1.0ms
Default Trigger Delays for 2-Wire and 4-Wire Resistance:
Range
Trigger Delay
(For NPLC
1)
Trigger Delay
(For NPLC <1)
100
1.5ms
1.0ms
1k
1.5ms
1.0ms
10k
1.5ms
1.0ms
100k
1.5ms
1.0ms
1M
1.5ms
10ms
10M
100ms
100ms
100M
100ms
100ms
Default Trigger Delays for AC Voltage and AC Current (all ranges):
AC Filter
Trigger Delay
3Hz - 300kHz filter (Slow)
7.0sec
20Hz - 300kHz filter (Medium)
1.0sec
200Hz - 300kHz filter (Fast)
600ms
Default Trigger Delay for Frequency and Period:
1.0s
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...