164
Multimeter Command Reference
*ESR?
*ESR?
returns the value of the Standard Event Status Register. The register is then
cleared (all bits 0).
Comments
•
Executable when Initiated:
Yes
•
Coupled Command:
No
•
*RST
Condition:
none
•
Power-On Condition:
register is cleared
*IDN?
*IDN?
returns identification information for the E1412A C-size multimeter. The
response consists of four fields:
HEWLETT-PACKARD,E1412A,0,A.0x.0x-A.0x.0x
The first two fields identify this instrument as model number E1412A manufactured
by Agilent Technologies (but listed as Hewlett-Packard). The third field is 0 since
the serial number of the multimeter is unknown to the firmware. The last field
indicates the revision level of the inguard-outguard firmware.
Note
The firmware revision field will change whenever the firmware is revised.
A.01.00-A.01.00 is the initial revision. The first two digits indicate the major
revision number and increment when functional changes are made. The last two
digits indicate the functional improvement level.
Comments
•
Executable when Initiated:
Yes
•
Coupled Command:
No
•
*RST
Condition:
none
•
Power-On Condition:
register is cleared
*OPC
*OPC
causes the E1412A to wait for all pending operations to complete after which
the Operation Complete bit (bit 0) in the Standard Event Status Register is set. The
*OPC
suspends any other activity on the bus until the multimeter completes all
commands sent to it prior to the
*OPC
command.
Comments
•
Executable when Initiated:
Yes
•
Coupled Command:
No
•
Related Commands:
*OPC?
,
*WAI
•
*RST
Condition:
none
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...