Agilent E1412A Multimeter Module Setup
15
Chapter 1
Agilent E1412A Multimeter
Module Setup
Using This Chapter
This chapter provides one page of general module information followed by
the tasks you must perform to set up your module and verify your
installation was successful. Chapter contents are:
•
Setting the Module Address Switch
•
Interrupt Priority
•
Setting and Checking the Line Frequency Reference
•
Input Terminals and Front Panel Indicators
•
Multimeter Functional Connections
•
Initial Operation
General Information
•
The E1412A (VXI C-size) Multimeter is a VXIbus message-based
slave device.
•
Programming the multimeter can either be through a command module
using an GPIB interface or an embedded controller. You use the
Standard Commands for Programmable Instruments (SCPI; see
Chapter 3) with the Standard Instrument Control Language (SICL) or
VISA (Virtual Instrument Software Architecture).
•
Maximum voltage is 300 V
rms
or 300 V
dc
.
•
Maximum current is 3A AC
rms
or DC.
•
Resolution is from 4½-digits for fast measurements to 6½-digits for
more accuracy. Resolution is set by specifying the integration time in
number of power line cycles (NPLCs) or corresponding aperture time.
Table 1-1 shows the correlation between NPLCs and resolution.
Table 1-1. Resolution of Power Line Cycles
Power Line Cycles
Resolution
0.02
0.0001 x Full-Scale
0.2
0.00001 x Full-Scale
1
0.000003 x Full-Scale
10
0.000001 x Full-Scale
100
0.0000003 x Full-Scale
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...