E1412A Multimeter Specifications
171
Appendix A
E1412A Multimeter Specifications
DC Characteristics
Accuracy Specifications
(% of r % of range) [1]
Function
Range
[3]
Test Current or
Burden
Voltage
24 Hour
[2]
23
C
1
C
90 Day
23
C
5
C
1 Year
23
C
5
C
Temperature
Coefficient
0
C - 18
C
28
C - 55
C
DC Voltage
100.0000mV
1.000000V
10.00000V
100.0000V
300.0000V
0.0030 + 0.0030
0.0020 + 0.0006
0.0015 + 0.0004
0.0020 + 0.0006
0.0020 + 0.0018
0.0040 + 0.0035
0.0030 + 0.0007
0.0020 + 0.0005
0.0035 + 0.0006
0.0035 + 0.0030
0.0050 + 0.0035
0.0040 + 0.0007
0.0035 + 0.0005
0.0045 + 0.0006
0.0045 + 0.0030
0.0005 + 0.0005
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0003
Resistance [4]
100.0000
1.000000k
10.00000k
100.0000k
1.000000M
10.00000M
100.0000M
1mA
1mA
100
A
10
A
5
A
500nA
500nA || 10M
0.0030 + 0.0030
0.0020 + 0.0005
0.0020 + 0.0005
0.0020 + 0.0005
0.0020 + 0.0010
0.0150 + 0.0010
0.3000 + 0.0100
0.0080 + 0.0040
0.0080 + 0.0010
0.0080 + 0.0010
0.0080 + 0.0010
0.0080 + 0.0010
0.0350 + 0.0010
0.8000 + 0.0100
0.0100 + 0.0040
0.0100 + 0.0010
0.0100 + 0.0010
0.0100 + 0.0010
0.0100 + 0.0010
0.0540 + 0.0010
0.8000 + 0.0100
0.0006 + 0.0005
0.0006 + 0.0001
0.0006 + 0.0001
0.0006 + 0.0001
0.0010 + 0.0002
0.0030 + 0.0004
0.1500 + 0.0002
DC Current
10.00000mA
100.0000mA
1.000000A
3.000000A
<0.1V
<0.7V
<1V
<2V
0.0050 + 0.0100
0.0100 + 0.0040
0.1000 + 0.0060
0.7000 + 0.0200
0.0500 + 0.0200
0.0500 + 0.0050
0.1300 + 0.0100
0.7200 + 0.0200
0.0700 + 0.0200
0.0700 + 0.0050
0.1500 + 0.0100
0.7200 + 0.0200
0.0050 + 0.0020
0.0060 + 0.0005
0.0060 + 0.0010
0.0060 + 0.0020
DC:DC Ratio
100mV to 300V
(Input Accuracy) + (Reference Accuracy)
Input Accuracy = accuracy specification for the HI-LO input signal.
Reference Accuracy = accuracy specification for HI-LO reference
input signal (Sense HI-LO input terminals).
NOTE:
Autorange is used for the reference signal regardless of the range set
for the HI-LO input signal. The 10V range is the highest range available
for the reference signal and the highest range the multimeter will
autorange to for measuring the reference signal.
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...