E1412A Multimeter Specifications
179
General Specifications
Overvoltage Category 1 (1500V peak max impulse)
E1412A Available Power (Amps):
+5V:
Ipm (maximum peak current):
0.20A
Idm (maximum dynamic current): 0.10A
+12V:
Ipm (maximum peak current):
0.70A
Idm (maximum dynamic current): 0.06A
Cooling /Slot:
Average Watts/Slot:
9.40
deltaP mm H2O:
0.05
Air Flow liters/s:
0.80
Operating Environment:
0
C to 55
C
65% Relative Humidity to 40
C
NOTE:
Recalibration may be required after exposure to humidity levels >65%.
Storage Environment:
-40
C to 70
C
State Storage Memory:
Power-off state automatically saved
Warm-up Time:
1 hour
Programming Language:
SCPI-1993, IEEE-488.2