Agilent E1412A Multimeter Error Messages
189
532
Cannot achieve requested resolution
The multimeter cannot achieve the requested measurement resolution. You
may have specified an invalid resolution in the
CONFigure
or
MEASure
command.
540
Cannot use overload as math reference
The multimeter cannot store an overload reading (9.9037) as the
math reference for null or dB measurements. The math state is turned off as
a result of this condition.
Self-Test Errors
The following errors indicate failures that may occur during a self-test.
The error message provides a description of the failure. Refer to the
E1412A
Service Manual
for more information.
602
RAM read/write failed
603
A/D sync stuck
604
A/D slope convergence failed
605
Cannot calibrate rundown gain
606
Rundown gain out of range
607
Rundown too noisy
609
DC gain x1 failed
610
DC gain x10 failed
611
DC gain x100 failed
612
Ohms 500nA source failed
613
Ohms 5µA source failed
614
DC 300V zero failed
615
Ohms 10µA source failed
616
DC current sense failed
617
Ohms 100µA source failed
618
DC high voltage attenuator failed
619
Ohms 1mA source failed
620
AC rms zero failed