194
Measurement Speed and Accuracy Trade-offs
Speed Advantage
Using the Special
Non-SCPI
Commands
(F1-F4 and R1-R7)
You can save approximately three (3) milliseconds by using an
F1
-
F4
special function command instead of changing function with the equivalent
SCPI
[SENSe:]
function command. You can save approximately five (5)
milliseconds by using an
R1
-
R7
special range command instead of
changing the range with the equivalent SCPI
[SENSe:]
range command.
Special Commands and Their Equivalent [SENSe:] Command
Speed
Advantage
Special
Command
Equivalent [SENSe:] SCPI command
approximate
3 mS advan-
tage with spe-
cial command
F1
[SENSe:]FUNC "VOLT[:DC]"
F2
[SENSe:]FUNC "CURR[:DC]"
F3
[SENSe:]FUNC "RES"
F4
[SENSe:]FUNC "FRES"
The special range command affects the currently configured function.
approximate
5 mS advan-
tage with spe-
cial command
R1
[SENSe:]VOLT[:DC]:RANG 0.1
[SENSe:]CURR[:DC]:RANG 0.01
[SENSe:]RES:RANG 100
[SENSe:]FRES:RANG 100
R2
[SENSe:]VOLT[:DC]:RANG 1
[SENSe:]CURR[:DC]:RANG 0.1
[SENSe:]RES:RANG 1000
[SENSe:]FRES:RANG 1000
R3
[SENSe:]VOLT[:DC]:RANG 10
[SENSe:]CURR[:DC]:RANG 1
[SENSe:]RES:RANG 10000
[SENSe:]FRES:RANG 10000
R4
[SENSe:]VOLT[:DC]:RANG 100
[SENSe:]CURR[:DC]:RANG 3
[SENSe:]RES:RANG 1E5
[SENSe:]FRES:RANG 1E5
R5
[SENSe:]VOLT[:DC]:RANG 300
CURR[:DC]
DOES NOT APPLY
[SENSe:]RES:RANG 1E6
[SENSe:]FRES:RANG 1E6
R6
VOLT[:DC]
DOES NOT APPLY
CURR[:DC]
DOES NOT APPLY
[SENSe:]RES:RANG 10E6
[SENSe:]FRES:RANG 10E6
R7
VOLT[:DC]
DOES NOT APPLY
CURR[:DC]
DOES NOT APPLY
[SENSe:]RES:RANG 100E6
[SENSe:]FRES:RANG 100E6
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...