Agilent E1412A Multimeter Application Information
25
Chapter 2
Agilent E1412A Multimeter Application
Information
Using This Chapter
This chapter provides multimeter application information in five parts.
•
Measurement Tutorial.
•
Measurement Configuration.
•
Math Operations.
•
Triggering the Multimeter.
•
E1412A Multimeter Application Examples.
Measurement Tutorial
The E1412A is capable of making highly accurate measurements. In order
to achieve the greatest accuracy, you must take the necessary steps to
eliminate potential measurement errors. This section describes common
errors found in measurements and gives suggestions to help you avoid these
errors.
DC Voltage Measurements
Thermal EMF
Errors
Thermoelectric voltages are the most common source of error in low-level
dc voltage measurements. Thermoelectric voltages are generated when you
make circuit connections using dissimilar metals at different temperatures.
Each metal-to-metal junction forms a
thermocouple
, which generates a
voltage proportional to the junction temperature. You should take the
necessary precautions to minimize thermocouple voltages and temperature
variations in low-level voltage measurements. The best connections are
formed using copper-to-copper crimped connections. Table 2-1 shows
common thermoelectric voltages for connections between dissimilar metals.
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...