Agilent E1412A Multimeter Application Information
37
Measurement Configuration
This section contains information to help you configure the multimeter for
making measurements. The parameters discussed in this section give you
measurement flexibility when using the
CONFigure
command.
AC Signal Filter
The E1412A Multimeter has three different ac filters which enable you to
either optimize low frequency accuracy or achieve faster ac settling times for
ac voltage or ac current measurements. Only these functions use the ac filter.
NOTE: These reading rates account for only the AC filters behavior.
See Page
for the effect of DC blocking circuitry.
•
The ac filter selection is stored in volatile memory. Default is the
medium filter (20Hz - 300kHz) at power-on or after a module reset.
•
The
CONFigure
and
MEASure:<
function
>?
commands automatically
select the medium (20Hz) filter.
•
Use the
[SENSe:]DETector:BANDwidth 3 | 20 | 200 | MIN | MAX
command to change the ac filter selection following a
CONFigure
command. The
MIN
parameter will select the 3Hz filter and the
MAX
parameter will select the 200Hz filter.
DC Input
Resistance
The E1412 Multimeter's input resistance is normally fixed at 10M
for all
dc voltage ranges to minimize noise pickup. You can set the input resistance
to greater than 10G
for the 100mVdc, 1Vdc and 10Vdc ranges to reduce
the effects of measurement loading errors. You select increased input
resistance using the
INPut:IMPedance:AUTO ON
command and this applies
to the dc voltage function only.
•
The input resistance setting is stored in volatile memory.
INPut:IMPedance:AUTO OFF
is set at power-on and after a module
reset.
•
The
CONFigure
command and the
MEASure:<
function
>?
command
automatically turn
AUTO OFF
. Use
INPut:IMPedance:AUTO ON
after a
CONFigure
command to set it
ON
.
Table 2-4. AC Signal Filters
AC Voltage or Current
Input Frequency
AC Filter
Selected
Max Reading Rate for
Adequate Settling
3 Hz to 300 kHz
Slow filter
1 reading/7 seconds
20 Hz to 300 kHz
Medium filter
1 reading/second
200 Hz to 300 kHz
Fast filter
10 readings/second
Table 2-5. DC Voltage Input Resistance
DC Input Resistance
100mV, 1V, 10V Ranges
DC Input Resistance
100V and 300V Ranges
INP:IMP:AUTO OFF
(DEFAULT)
10M
10M
INP:IMP:AUTO ON
>10G
10MW
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...