Agilent E1412A Multimeter Application Information
51
Querying the
Delay Time
The
TRIGger:DELay? [MINimum|MAXimum]
command returns one of the
following numbers to the output buffer:
•
The present trigger delay (1
s through 3600 seconds) if
MIN
or
MAX
is
not specified.
•
The minimum trigger delay available (1
s) if
MIN
is specified.
•
The maximum trigger delay available (3600 seconds) if
MAX
is
specified.
The Sample Count
The
SAMPle:COUNt <
number
>
command designates the number of
readings per trigger. The
number
parameter sets the number of readings to
a value between 1 and 50,000.
Substituting
MIN
for the
number
parameter sets the number of readings per
trigger to 1. Substituting
MAX
for the
number
parameter sets the number of
readings per trigger to 50,000.
Example: Setting the
Sample Count
In the following example, 10 DC voltage measurements are made when the
multimeter's external trigger BNC connector is pulsed low. After the
10 readings are taken, the multimeter returns to the idle state.
dimension array
Dimension computer array.
CONF:VOLT:DC
Function: DC voltage.
TRIG:SOUR EXT
Trigger source is external BNC on multimeter front
panel.
SAMP:COUN 10
Specify 10 readings per trigger.
READ?
Place multimeter in wait-for-trigger state; make
measurements when external trigger is received; send
readings to output buffer.
timeout may occur
May require INIT, monitor the status byte for
completion (standard event bit 0), FETC? to transfer
readings to the output buffer (vs. READ?).
enter statement
Enter readings into computer.
Checking the
Sample Count
The
SAMPle:COUNt? [MINimum|MAXimum]
command returns one of the
following numbers to the output buffer:
•
The present sample count (1 through 50,000) if neither
MIN
nor
MAX
is specified.
•
The minimum sample count available (1) if
MIN
is specified.
•
The maximum sample count available (50,000) if
MAX
is specified.
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...