72
Multimeter Command Reference
ABORt
The
ABORt
command subsystem removes the multimeter from the wait-for-trigger
state and places it in the idle state.
ABORt
is only effective when the trigger source
is
TRIGger:SOURce BUS
.
Subsystem Syntax
ABORt
Example
Aborting a Measurement
CONF:VOLT:DC
Function: DC voltage.
TRIG:SOUR BUS
Trigger source is BUS trigger.
INIT
Place multimeter in wait-for-trigger state.
ABOR
Abort waiting for a trigger and place
multimeter in idle state.
Comments
•
ABORt
does not affect any other settings of the trigger system. When the
INITiate
command is sent, the trigger system will respond as it did before
ABORt
was executed.
•
ABORt
returns the multimeter to the idle state for
TRIGger:SOURce BUS
. The
“Trigger ignored” error is generated when a Group Execute Trigger (GET) bus
command or
*TRG
common command is executed after an
ABORt
command
(which puts the multimeter into the idle state).
•
Related Commands:
INITiate
,
TRIGger
•
*RST
Condition:
After a a
*RST
, the multimeter acts as though an
ABORt
has
occurred.