Multimeter Command Reference
:NULL:OFFSet
CALCulate:NULL:OFFSet <
value
>|MIN|MAX
stores a null value in the
multimeter's Null Register.
You must turn on the math operation e.g., execute
CALC:STAT ON before writing to the math register.
You can set the null value to
any number between 0 and
120% of the highest range, for the present function.
MIN
= –120% of the highest range.
MAX
= 120% of the highest range. The null value
is stored in volatile memory. See the section titled “Math Operations - NULL
Function” beginning on page 41 for another way to store the offset value.
Example
Set the Null Offset Value
CALC:FUNC NULL
Set math function to NULL. You may choose to
set the math function after setting STATe ON.
CALC:STAT ON
Turn on math operation.
CALC:NULL:OFFS 500
Set null offset to 500.
:NULL:OFFSet?
CALCulate:NULL:OFFSet? [MIN|MAX]
queries the null value.
Example
Query the Null Offset Value Set for the
NULL
Math Operation
CALC:NULL:OFFS?
Query the null offset value.
:STATe
CALCulate:STATe OFF|ON
disables or enables the selected math function. The
state is stored in volatile memory.
Example
Enable the Currently Selected Calculate Math Function
CALC:STAT ON
The selected or default math function is
enabled.
:STATe?
CALCulate:STATe?
queries the state of the math function. Returns “
0
” (
OFF
) or
“
1
” (
ON
).
Example
Query Whether a Math Function State is On or Off
CALC:STAT?
Query the state.
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...