80
Multimeter Command Reference
:LFRequency?
CALibration:LFRequency?
queries the line frequency setting.
Comments
This command returns
+50
for line frequency set to 400 because 400 is an even
multiple of 50.
Example
Query the Line Frequency Setting
CAL:LFR?
Query the line frequency.
:SECure:CODE
CALibration:SECure:CODE
<
new code
>
enters a new calibration security code.
To change the security code, first unsecure the multimeter using the old security code
with
CAL:SEC:STAT OFF, <
old code
>
. Then, enter the new code. The calibration
security code may contain up to 12 characters. The security code is stored in
non-volatile memory.
Comments
•
The security code is set to “HP_
E1412
” for C-size when the multimeter is
shipped from the factory. The security code is stored in non-volatile memory,
and does not change when power has been off or after a remote interface reset.
•
The security code
<
new code
>
can contain up to 12 alphanumeric characters.
The first character must be a letter. The remaining characters can be letters or
numbers or an underscore. You do not have to use all 12 characters but the first
character must be a letter.
•
If you forget or lose the active security code, you can disable the security
feature by adding a jumper inside the multimeter (see Chapter 5 in the Service
Manual). You then enter a new code and remove the jumper.
Example
Enter a New Calibration Security Code
CAL:SEC:STAT OFF, HP_E1412
Unsecure with the old code.
CAL:SEC:CODE
the_new_code
Enter a new calibration code
(a maximum of 12 characters).
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...