Multimeter Command Reference
83
:ZERO:AUTO
CALibrate:ZERO:AUTO <
mode
>
enables or disables the autozero mode. Autozero
applies to dc voltage, dc current and 2-wire ohms measurements only. 4-wire ohms
and dc voltage ratio measurements automatically enable the autozero mode.
Parameters
Comments
•
You can use “
0
” for
OFF
and “
1
” for
ON
in the mode parameter.
•
The
ON
parameter enables autozero. This is the default parameter which
causes the multimeter to internally disconnect the input signal following each
measurement and make a zero measurement. The zero reading is subtracted
from the input signal reading to prevent offset voltages present on the
multimeter's input circuitry from affecting measurement accuracy.
•
The
OFF
parameter disables autozero. In this mode the multimeter takes one
zero measurement and subtracts it from all subsequent input signal
measurements prior to a change in function, range or integration time. A new
zero measurement is made following a change in function, range or integration
time. This mode increases measurement speed because a zero measurement is
not made for each input signal measurement.
•
Autozero
ONCE
issues an immediate zero measurement and can be used to get
an update on the zero measurement for a specific input signal measurement.
This helps to increase measurement speed since you update the zero reading
without making zero measurements for every measurement.
•
*RST
Condition:
CALibrate:ZERO:AUTO ON
(autozero enabled)
:ZERO:AUTO?
CALibrate:ZERO:AUTO?
queries the autozero mode. Returns “
0
” (
OFF
or
ONCE
)
or “
1
”
ON
.
Parameter Name
Parameter Type
Range of Values
Default Units
<
mode
>
boolean
OFF|0|ON|1|ONCE
none
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...