:FRESistance
CONFigure:FRESistance [<
range
>|MIN|MAX|DEF|AUTO
[,<
resolution
>|MIN|MAX|DEF]]
selects the 4-wire ohms function and allows you to
specify the measurement range and resolution.
Parameters
Comments
•
To select a standard measurement range, specify
range
as the input signal's
maximum expected resistance. The multimeter then selects the correct range
that will accept the input.
•
The
AUTO
or
DEFault
option for the
range
parameter enables autorange. The
DEFault
option for
resolution
defaults the integration time to 10 PLC.
•
The
MIN
and
MAX
parameters select the minimum or maximum values for
range
and
resolution
:
For
range
:
MIN
= 100
;
MAX
=100M
For
resolution
:
MIN
selects the best resolution (the smallest value) for the
selected range.
MAX
selects the worst resolution (the largest value) for the
selected range.
•
To select autorange, specify
DEF
for
range
or do not specify a value for the
range and resolution parameters. In the autorange mode, the multimeter
samples the input signal before each measurement and selects the appropriate
range.
•
To specify a
MIN
or
MAX
resolution while autoranging, you must
specify the
AUTO
or
DEFault
parameter;
CONF:FRES DEF,MIN
or
CONF:FRES DEF,MAX
or
CONF:FRES AUTO,MIN
or
CONF:FRES AUTO,MAX
(you cannot omit the
range
parameter). This
prevents the
MIN
or
MAX
resolution from being interpreted as a range setting.
•
Related Commands:
FETCh?
,
INITiate
,
READ?
Example
Making 4-Wire Ohms Measurements
CONF:FRES 1500,MAX
Function: 4-wire ohms; range selected: 10 k
;
M
AX resolution: 1
SAMP:COUN 3
Take 3 readings; trigger source is IMMediate
by default.
READ?
Place multimeter in wait-for-trigger state and
make measurements; send readings to output
buffer.
enter statement
Enter readings into computer.
Parameter Name
Parameter Type
Range of Values
Default Units
<
range
>
numeric
100
|1k
|10k
|100k
|1M
|10M
|100M
|
MIN|MAX|DEF|AUTO
ohms
<
resolution
>
numeric
resolution
|MIN|MAX|DEF
ohms
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...