Multimeter Command Reference
CONFigure?
The
CONFigure?
command queries the multimeter to return the configuration set by
the most recent
CONFigure
or
MEASure
command.
It returns a quoted string to the output buffer in the following format:
“<
function
> <
parameter
>,<
parameter
>”
Subsystem Syntax
CONFigure?
Comments
•
When the multimeter is configured for current, voltage or resistance
measurements,
CONFigure?
returns the function followed by the selected
range and resolution. For example:
“CURR:AC +1.00,1.000000E-05”
“CURR +1.00,1.000000E-05”
“VOLT:AC +2.02,1.000000E-06”
“VOLT +3.02,1.000000E-06”
“FRES +100.0000E+03,1.000000E-05”
“RES +1.03,1.000000E-03”
“FREQ +3.00,3.000000E-05”
“PER +3.333330E-01,3.333330E-06”
•
If you specify
DEF
,
MIN
, or
MAX
for the
range
or
resolution
parameters in
CONFigure
or
MEASure
, the
CONFigure?
command returns the selected
value.
•
Related Commands:
CONFigure
,
MEASure
Example
Querying the Multimeter Configuration
dimension string array
Dimension computer array to store string.
CONF:FRES 900,MAX
Function: 4-wire ohms; range selected: 1k
;
MAX resolution: 100m
.
CONF?
Query configuration.
enter statement
Enter string into computer.
String Returned:
“FRES +1.003,9.999999E-02”
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...