Chapter 4
127
Troubleshooting Information
Checking PC Client Connections
viopen
fails or I/O
operation fails: Out
of resources
The E5810 was unable to open another session or perform the
operation because it is out of resources. You may need to have a client
terminate its connection or you may need to reboot the E5810 to close
all client connections to reclaim resources. This action will disconnect
all other users accessing the E5810.
I/O operation fails:
No connection
The connection to the E5810 failed because:
The network connection is down or the E5810 is down and a
network timeout occurred.
The E5810 was powered off and then on or an E5810 reboot
was executed, either of which terminates any open client
connections.
I/O operation fails:
Timeout
There are several possible causes:
The client may be attempting a transaction to a non-existent or
powered-off device. Check the address used, as well as the
status of the device.
The network connection or the E5810 has gone down since the
previous I/O operation.
The I/O timeout specified is not long enough for the transaction to
complete. You may need to verify that the E5810 I/O timeout
configuration value is long enough if the client specified a timeout
value of infinity.
The E5810 is busy performing an I/O operation for another client
or another client has the device/interface locked. See “E5810
appears to be hung” for more information.
Message
Description/Actions
Summary of Contents for Agilent E5810A
Page 1: ...Agilent E5810A LAN GPIB Gateway for Windows User s Guide ...
Page 2: ......
Page 6: ...Contents 6 ...
Page 15: ...1 E5810 Description ...
Page 33: ...2 Installing the E5810 ...
Page 75: ...3 Using E5810 Web Access ...
Page 85: ...Chapter 3 85 Using E5810 Web Access Viewing and Modifying Configuration ...
Page 108: ...108 Chapter 3 Using E5810 Web Access Other Web Access Functions ...
Page 109: ...4 Troubleshooting Information ...
Page 131: ...A E5810 Specifications ...
Page 136: ...136 AppendixA E5810 Specifications Supported Network Protocols Notes ...
Page 137: ...B Using the Agilent IO Libraries ...
Page 152: ...152 Appendix B Using the Agilent IO Libraries Installing Custom Agilent IO Libraries Notes ...
Page 153: ...C Using the Telnet Utility ...
Page 160: ...160 Appendix C Using the Telnet Utility Telnet Commands for the E5810 Notes ...
Page 161: ...161 Glossary ...
Page 172: ...172 Glossary Glossary Notes ...
Page 176: ...176 Index ...