Chapter 4
Troubleshooting Information
Checking Instruments
Checking RS-232 Instruments
If you cannot establish communication with a connected RS-232 instrument,
see the following table for suggested troubleshooting steps.
Possible Cause(s)
Corrective Action(s)
Attempting to Communicate From the E5810 Web Access
Intermittent or missing RS-232
connection
Check RS-232 cable connections.
Verify that the RS-232 cable is the
correct cable for this application.
RS-232 device power not ON or
defective RS-232 device
Make sure RS-232 device power is
ON and the device is operating
properly.
RS-232 interface not configured for
attached device
Change E5810 RS-232 interface
attributes to match RS-232 device
attributes (baud rate, parity, etc.).
Attempting to Communicate Programatically
VISA/SICL not configured
Run the IO Config Utility (see
Chapter 2 - Installing the E5810
).
Incorrect VISA/SICL address
Check VISA/SICL address being
used.
Summary of Contents for Agilent E5810A
Page 1: ...Agilent E5810A LAN GPIB Gateway for Windows User s Guide ...
Page 2: ......
Page 6: ...Contents 6 ...
Page 15: ...1 E5810 Description ...
Page 33: ...2 Installing the E5810 ...
Page 75: ...3 Using E5810 Web Access ...
Page 85: ...Chapter 3 85 Using E5810 Web Access Viewing and Modifying Configuration ...
Page 108: ...108 Chapter 3 Using E5810 Web Access Other Web Access Functions ...
Page 109: ...4 Troubleshooting Information ...
Page 131: ...A E5810 Specifications ...
Page 136: ...136 AppendixA E5810 Specifications Supported Network Protocols Notes ...
Page 137: ...B Using the Agilent IO Libraries ...
Page 152: ...152 Appendix B Using the Agilent IO Libraries Installing Custom Agilent IO Libraries Notes ...
Page 153: ...C Using the Telnet Utility ...
Page 160: ...160 Appendix C Using the Telnet Utility Telnet Commands for the E5810 Notes ...
Page 161: ...161 Glossary ...
Page 172: ...172 Glossary Glossary Notes ...
Page 176: ...176 Index ...