4263B
LCR Meter
LCR & Resistance Meters
4263B LCR Meter
The 4263B LCR meter is Agilent Technologies’ most cost-effective
low-end LCR meter, designed for both component evaluation on the
production line and fundamental impedance testing for bench-top
applications. The 4263B has five test frequencies that allow you to
simulate testing under the correct conditions: 100 Hz, 120 Hz, 1 kHz,
10 kHz, and 100 kHz. An optional 20 kHz test frequency can be
added to those five frequencies (4263B-002).
High-Speed Measurements
The 4263B can boost throughput with a measurement speed of 25 ms
at any test frequency. This ability improves the throughput of elec-
trolytic capacitor and transformer testing. The 4263B can check the
contact condition between the test terminals and the device-under-
test (DUT). This function ensures the reliability of PASS/FAIL testing
with automatic handlers in production. The quick recovery system of
the 4263B improves throughput. Normal operation is resumed the
instant a faulty DUT is removed from the handler, so the handler can
always be operated at its full speed.
Electrolytic Capacitor Measurements
The 4263B’s accuracy and wide measurement range are the right
tools to make precise measurements of electrolytic capacitors.
Charged capacitors can discharge through the front end and destroy
an instrument. The 4263B’s front end is designed for protection and
maintains test integrity.
Transformer Parameter Measurements
With the 4263B’s ability to make turns ratio (N), mutual inductance
(M), and dc resistance (DCR) measurements, data calculations and
changing test setups are no longer time-consuming tasks (4263B-001).
The flexible signal level setting and the voltage-and-current monitor
function facilitate the use of the 4263B for level dependent DUTs, such
as core inductors.
Specifications
(Refer to Technical Overview for complete specifications)
Measurement Functions
• Measurement Parameters: |Z|, |Y|, u, R, X, G, B, L, C, Q, D, ESR
– 4263B-001: Add DCR (dc resistance), N (turns ratio),
and M (mutual inductance) measurement
• Measurement Circuit Mode: Series and parallel
Mathematical Functions:
Deviation and percent deviation
Test Cable Lengths:
0 m, 1 m, 2 m, 4 m (freq. = 100/120/1 kHz);
0 m,1 m, 2 m (freq. = 10 k/20 kHz); 0 m, 1 m (freq. = 100 kHz)
Test Signal Information
• Test Frequency: 100 Hz, 120 Hz, 1 kHz, 10 kHz, and 100 kHz
– 4263B-002: Add 20 kHz test frequency
• Frequency Accuracy: ±0.01% (freq. = 100 Hz, 1 kHz, 10 kHz,
(20 kHz, 100 kHz), ±1% (freq. = 120 Hz)
• Output Impedance: 100
Ω
±10%, 25
Ω
±10% (
≤
1
Ω
range)
• AC Test Signal Level: 20 mV to 1 V
rms
in 5 mV
rms
steps
• Accuracy: ±(10% + 10 mV)
• Internal dc Bias
– Level: 1.5 and 2 V; Accuracy: ±(5% + 2 mV)
• External dc Bias: 0 V to +2.5 V
Measurement Range
Parameter
Measurement Range
Z
, R, X
1 m
Ω
to 100 M
Ω
Y
, G, B
10 nS to 1000 S
C
1 pF to 1 F
L
10 nH to 100 kH
D
0.0001 to 9.9999
Q
0.1 to 9999.9
–180° to +180°
DCR
1 m
Ω
to 100 M
Ω
N
0.9 to 200 (unspecified)
L, M
1 µH to 100 H (unspecified)
%
–999.99% to +999.99%
Measurement Accuracy:
±0.1% (basic) (for
Z
, R, X,
Y
, G, B,C, L)
Measurement Time
Mode
Time (typical)
SHORT
25 ms
MEDIUM
65 ms
LONG
500 ms
Test Signal Level Monitor:
Voltage and current
Front-End Protection:
Internal circuit protection when a charged capacitor
is connected to the input terminals. The maximum capacitor voltage is:
Vmax =
√
(8/C) (typical) @ Vmax
≤
250 V; Vmax =
√
(2/C) (typical) @ Vmax
≤
1000 V, C is in Farads
Display Digits:
3, 4, or 5 (selectable)
Correction Function
OPEN/SHORT/LOAD: Eliminates measurement errors due to stray
parasitic impedances in the test fixtures
Comparator Function:
HIGH/IN/LOW for each primary measurement
parameter and secondary measurement parameter
Contact Check Function:
Contact failure between the test fixture and
device can be detected. Additional time for contact check: 5 ms
Other Functions
• Save/Recall: Ten instrument setups can be saved/recalled from
the internal nonvolatile memory
• Continuous Memory Capability: If the instrument is turned off,
or if a power failure occurs, instrument settings (except dc bias
on/off) are automatically memorized (
≤
72 hours at 23°C ± 5°C)
• GPIB Interface: All control settings, measured values, and
comparator information
• Handler Interface: All output signals are negative-logic, optically
isolated open collectors. Output signals include HIGH/IN/LOW,
no contact, index, end of measurement, and alarm. Input signals
include keylock and external trigger
General Specifications
Power Requirements:
90 to 132 V or 198 to 264 V, 47 to 66 Hz, 45 VA max.
Operating Temperature:
0°C to 45°C
Size:
100 mm H x 320 mm W x 300 mm D (3.94 in x 12.6 in x 11.81 in)
Weight:
4.5 kg (9.9 lb)
Key Literature
4263B LCR Meter Technical Overview, p/n 5964-6181E
Effective Electrolytic Capacitors Testing, p/n 5967-5378E
Effective Transformer/LF Coil Testing, p/n 5967-5377E
Effective Multi-tap Transformer Measurement using a Scanner and the
Agilent 4263B LCR Meter, p/n 5091-6310E
Ordering Information
4263B
LCR Meter
Options
4263B-001
Add Transformer Parameter Measurement Function
4263B-002
Add 20 kHz Test Frequency
4263B-ABA
US-English Localization
4263B-ABJ
Japan-Japanese Localization
Accessories
16060A
Transformer Test Fixture (4263B-001 Required)
16065C
External Bias Adapter (up to 40 Vdc)
16034G
Small SMD Test Fixture (0603 [mm]/0201 [in] SMD)
16044A
Kelvin Contact SMD Test Fixture
16047A
4-Terminal Contact Lead Component Test Fixture
16047E
2-Terminal Contact Lead Component Test Fixture
16334A
Tweezers Contact SMD Test Fixture
• 0.1% basic accuracy
• 100 Hz, 120 Hz, 1 kHz, 10 kHz, 100 kHz test frequencies
• 20 mV to 1 V
rms
in 5 mV
rms
steps
• Test signal level monitor function
• High-speed measurement: 25 ms
• High-speed contact check
• Wide capacitance test range
• Transformer parameter measurements (optional)
4263B