3 Common AssayMAP Bravo error messages
Robot disable is active messages
12
AssayMAP Bravo Platform Error Recovery Guide
Restoring the Bravo device after an emergency stop
To restore the Bravo device after an emergency stop:
1
At the emergency-stop pendant, turn the red button clockwise. The spring-loaded
button pops up.
2
If applicable
. Remove any object that is interrupting the Light Curtain.
3
In the
AssayMAP Bravo Error
dialog box, click
Retry
.
In most cases, the AssayMAP Bravo Platform will be able to resume the run where
it left off.
Note:
In some rare cases, the
W-Axis commanded to invalid destination
error message
may appear after attempting to retry. If this happens, see the
to an invalid destination ” on page 13
for instructions on how to recover.