3 Common AssayMAP Bravo error messages
W-axis was commanded to an invalid destination
AssayMAP Bravo Platform Error Recovery Guide
W-axis was commanded to an invalid destination
Error message example
In some rare cases, when the interlock circuit is tripped, the
W-axis was commanded to
an invalid destination
error message appears. The error message provides information
about the deck location and the task that the AssayMAP Bravo Platform was
performing when the error occurred.
Recovery procedure
The correct recovery procedure depends on the task in progress at the time and the
deck location. The error message contains these details. Use the following scenarios to
determine the correct recovery path for your situation:
•
Scenario 1
. The message occurs during a syringe wash at the wash station. See
“Syringe wash using the wash station at deck location 1” on page 14
.
•
Scenario 2
. The message occurs during an aspirate or dispense task that does not
involve the wash station at deck location 1. In this case, the interrupted run cannot
be recovered and must be aborted before starting a new run. See
Diagnostics to reclaim reagents from syringes” on page 18
.