3 Common AssayMAP Bravo error messages
W-axis was commanded to an invalid destination
14
AssayMAP Bravo Platform Error Recovery Guide
Syringe wash using the wash station at deck location 1
To recover from the error in this scenario:
1
If necessary, turn the red button clockwise on the emergency-stop pendant to reset
it, and ensure that no object is interrupting the Light Curtain.
2
In the
W-axis commanded to an invalid destination
error message, click
Retry
to
recover from the error and continue with the syringe washing process.