Introduction
Cary 50 Spectrophotometer Hardware Manual
13
Environmental
Table 1.
Suitable
conditions during instrument transportation, non-operation and operation
Condition
Altitude (m, ft)
Temperature
(°C, °F)
Relative humidity,
non-condensing (%)
Non-operating (transport)
0–3050, 0–10,000
5–45, 41–113
20–80
Operating within performance
specifications
0–3050, 0–10,000
10–35, 50–95
8–80
NOTE
For
optimum analytical performance
, it is recommended that the ambient
temperature of the laboratory be between
20 and 25 °C
and be held constant to
within
±
2 °C throughout the entire working day.
Weights and dimensions
Weights
Packed: 23 kg (50.7 lb)
Unpacked: 21 kg (46 lb)
Dimensions (width x depth x height)
Packed: 615 x 710 x 360 mm (24 x 28 x 14 in)
Unpacked: 590 x 500 x 200 mm (23 x 20 x 8 in)
Computer power supply
To ensure safe operation of the instrument, the switching power
supply in the computer must comply with standard IEC 950 or
equivalent.