Installation
Cary 50 Spectrophotometer Hardware Manual
17
2
If the cell aperture is not at the correct height, remove the cell
and adjust the grub screw in the microcell holder accordingly
using the 2.5 millimeter hexagonal ball driver. Replace the cell in
the microcell holder.
3
Slide back the sample compartment lid (refer to Figure 2).
1
2
Figure 2.
Cary 50 showing the sample compartment lid (1) and front panel (2)
4
Remove the sample compartment front panel (refer to Figure 2).
5
If not already fitted, install the cell holder base in the sample
compartment as follows:
a
Place the cell holder base in the sample compartment,
aligning the two locating holes over the two locating pins in
the floor of the sample compartment.
b
Firmly tighten the thumbscrew on the cell holder base (item
2 in Figure 1).
6
Place the microcell holder on the cell holder base, aligning the
holes in the microcell holder over the raised black knobs in the
cell holder base.
7
Tighten the thumbscrew on the microcell holder.
You should now align the microcell holder.