Maintenance
Cary 50 Spectrophotometer Hardware Manual
25
Aligning the source mirror
For optimum performance of the instrument, the source mirror
should be aligned to suit the new lamp module.
To align the source mirror:
1
Start the Align application by clicking the Windows
Start
button
and choosing
(All) Programs
,
Agilent
, Cary
WinUV
, (or
Agilent
,
UV/UV FO Dissolution
) then
Align
.
2
On the
Cary
tabbed page, set the following parameters:
Beam Mode: Single Beam, Normal
Y Mode: %T
Ave. Time: 0.5000
Wavelength: 500 nm
3
Click
Apply
.
4
Remove the snap-out panel (refer to Figure 5) from the front of
the instrument, using a flat-blade screwdriver to pry the panel
open. This exposes the two source mirror adjustment screws.
NOTE
Do not remove the plastic bung next to the adjustment screws.