40 RF Multiplexer Command Reference
Chapter 4
[ROUTe:]
The ROUTe command subsystem controls switching operations for the
RF Multiplexer in a switchbox.
Subsystem Syntax
[ROUTe:]
CLOSe
<channel_list>
CLOSe?
<channel_list>
OPEN?
<channel_list>
[ROUTe:]CLOSe
ROUTe:]CLOSe
<channel_list >
closes the RF Multiplexer channels specified
by
channel_list
.
channel_list
has the form (@
ccmmnn
) where
cc
= card number (01-99),
mm
= module number (00-02), and
nn
= channel number (00-03, 10-13, 20-23, 30-33, 40-43, or 50-53).
Parameters
Comments
Closing Channels:
To close a single channel, use [ROUTe:]CLOSe
(@
ccmmnn
); for multiple channels use [ROUTe:]CLOSe
(@
ccmmnn
,
ccmmnn
,...) or any combination. Closure order for multiple
channels with a single command is not guaranteed.
Channel Range:
The RF Multiplexer will accept and execute channel ranges
(
ccmmnn
:
ccmmnn
) without generating an error, but the result is to close the
last channel in each bank within the range specified. For example, after
CLOSe (@010101:010151) is executed, card 01, module 01 channels 03,
13, 23, 33, 43, and 51 would remain closed.
Module Number:
The module number can be omitted when Expander
modules are not connected to the RF Multiplexer. Module numbers must be
specified when one or more Expander modules are connected to the RF
Multiplexer module.
Related Commands:
[ROUTe:]OPEN?, [ROUTe:]CLOSe?
*RST Condition:
All banks with channel 0 connected to COMmon.
Name
Type
Range
Item
<channel_list>
numeric
numeric
numeric
01-99
00-02
00-03 |10-13 |20-23 |30-33 |40-43 |50-53
card (cc)
module (mm)
channel (nn)