44 RF Multiplexer Command Reference
Chapter 4
Comments
RF Multiplexer/Expander modules installed:
The SYSTem:COPTion?
<number>
command returns:
E1472A,E1473A,E1473A or E1474A,E1475A,E1475A
where the E1472A and E1474A are the RF Multiplexer modules, and the
E1473A,E1473A or E1475A,E1475 indicates that two expander modules
are connected. A “0” in place of the “E1473” or “E1475” indicates that the
Expander module(s) are not installed.
Example
Reading the Model Number of Card #1 Module
SYST:COPT? 1
!Return the module configuration
!for card 1
SYSTem:CPON
SYSTem:CPON <number> |ALL
sets the selected module (card) in a
switchbox to its power-on state.
Parameters
Comments
RF Multiplexer Module Power-On State:
The power-on state is all channels
(relays) open. Note that SYSTem:CPON ALL and *RST connects X0 to
COM X0 in all banks of all modules in a switchbox, while SYSTem:CPON
<number>
connects X0 to COMX0 in only the module (card) specified in the
command.
Example
Setting Card #1 Module to Power-On State
SYST:CPON 1
!Sets card #1 to power-on state
SYSTem:CTYPe?
SYSTem:CTYPe?
<number>
returns the card type of a selected module in a
switchbox.
Parameters
Name
Type
Range of Values
Default Value
<number>
numeric
01 to 99 | ALL
ALL
Name
Type
Range of Values
Default Value
<number>
numeric
01 to 99
N/A
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com