25
Hardware Installation
SI-9250 Cabling
Cabling for the SI-9250-2 tuner is shown in
.
Figure 13.
SI-9250 cabling
Special requirements:
• A 50
Ω
termination must be installed on the 100 MHz Reference output.
• A 10 MHz, 0 dBm signal must be applied to the 10 MHz Reference input
The configuration file entries for this tuner are as follows:
searchRx1.tuner1.tunerModel: SI9250
searchRx1.tuner1.tunerInterfaceParm: 144, 0, 10, 136, 0, 3
where:
144 = logical address of SI-9250
0 = Reference input: 0 = int or ext 10MHz, 1 = ext 100MHz
10 = SI-9250 settling time in ms
136 = logical address of companion tuner, E2730/31B
0 = 10MHz Ref input: 0 = internal, 1 = external
3 = E273x settling time in milliseconds, typically 3 to 5 ms
For directions on setting the address, see