Summary of Contents for E4402B Option H3B

Page 1: ...Agilent Technologies E4402B Option H3B User s and Service Guide ...

Page 2: ......

Page 3: ...ologies E4402B Option H3B User s and Service Guide Use this manual with the following document ESA E Manual Set Manufacturing Part Number E4401 90511 Printed in USA March 2006 Copyright Agilent Technologies Inc 2006 ...

Page 4: ...ENT THAT CONFLICT WITH THESE TERMS THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL DFARS Restricted Rights Notice If software is for use in the performance of a U S Government prime contract or subcontract Software is delivered and licensed as Commercial computer software as defined in DFAR 252 227 7014 June 1995 or as a commercial item as defined in FAR 2 101 a or as Restricted computer...

Page 5: ... understood and met Definitions Specifications describe the performance of parameters covered by the product warranty temperature 0 to 55 C unless otherwise noted Typical describes additional product performance information that is not covered by the product warranty It is performance beyond specification that 80 of the units exhibit with a 95 confidence level over the temperature range 20 to 30 C...

Page 6: ...iv ...

Page 7: ...epair Information 9 Replaceable Parts 10 Safety and Regulatory Information 11 Introduction 11 Before Applying Power 11 Connector Care and Cleaning 11 Declaration of Conformity 11 Statement of Compliance 11 Shipping Instructions 11 Compliance with Canadian EMC Requirements 12 Compliance with German Noise Requirements 12 Warnings 13 Cautions 14 Instrument Markings 15 Contacting Agilent Sales and Ser...

Page 8: ...Contents 2 Contents ...

Page 9: ...User s and Service Guide 1 E4402B Option H3B ...

Page 10: ...ut is greater than 5 dBm from 3 9214 GHz to 4 9214 GHz when the E4402B is tuned to a center frequency between 0 Hz and 1 0 GHz and is greater than 10 dBm for higher LO frequencies Center Frequencies greater than 1 GHz when the E4402B is tuned to a center frequency above 1 0 GHz Any device connected to this output must have a 50 Ω impedance The Agilent Technologies E4402B Option H3B comes with a 50...

Page 11: ...p the original packaging materials so they can be used if you need to transport the instrument Inspect the instrument and all accessories for any signs of damage that may have occurred during shipment If your instrument or any accessories appear to be damaged or missing refer to Contacting Agilent Sales and Service Offices on page 16 ...

Page 12: ...e Video In on the transmitter or modulator under test 3 Secure the appropriate adapters Type N to BNC to the ends of the BNC cable 8120 2582 and attach to the 50 Ω RF Input of the E4402B Spectrum Analyzer and the other end to the output of the modulator CAUTION A directional coupler or high power attenuator may be needed to keep the DUT output power from exceeding 30 dBm or 1 W max safe input WARN...

Page 13: ...re used throughout this document HARDKEYS are labeled front panel keys SOFTKEYS are unlabeled key whose function is indicated on the instrument display Modulator or Transmitter under test Aerodynex Sideband Adapter LO IN rear VIDEO OUT rear 50Ω RF INPUT LO OUT rear VIDEO IN RF OUTPUT E4402B with Option H3B 3 9214 GHz 4 9214 GHz High Power Attenuator ...

Page 14: ...h of the instrument operating manuals 2 Turn Off the Auto Align routine before making your measurements Perform the following keystrokes System Alignments Auto Align Off 3 Select the desired channel on the modulator or use the assigned transmitter frequency Channel two 55 25 MHz has been selected for this example 4 Perform the following keystrokes Frequency Start Freq 52 MHz Frequency Stop Freq 63...

Page 15: ...the desired frequency Use the Fine adjust knob if more accuracy is desired The waveform displayed on the E4402B Spectrum Analyzer will display the bandwidth of the transmitter or modulator under test Refer to Figure 3 Refer to the Aerodynex Sideband Adapter documentation for more information on making measurements Figure 3 Example of a Video Signal ...

Page 16: ...trum analyzer using an RF cable and adapter if required 4 Press Frequency on the ESA DUT and set the start frequency to 0 Hz and the stop frequency to 1 1 GHz Set the Sweep time to 10 sec 5 Press Frequency on the test spectrum analyzer and set the start frequency to 3 GHz and the stop frequency to 5 GHz 6 Set the RF level on the test spectrum analyzer to 0 dBm 7 Increase the test spectrum analyzer...

Page 17: ...ontains the Option H3B RF cable Verify that Option H3B has been installed by checking the serial tag on the rear panel of the instrument and verifying the terminated 3 5mm LO Output connector is installed on the rear panel of the instrument The following photos show the replaceable assemblies Refer to Replaceable Parts on page 10 Figure 4 LO Output J15 Figure 5 Cable Routing ...

Page 18: ...er so long lead times may be encountered when ordering replacement parts Description Agilent Part Number Qty Spiral Wrap 16 inches 0890 0025 1 Adapter Coax SMA f SMA f 1250 1666 1 Termination 50 Ω 1810 0118 1 Nut Hex 1 4 2950 0223 1 Washer 1 4 3050 0420 1 Cable Assembly 18 inches 5062 6665 1 Rear Panel H3B E4401 00085 1 ...

Page 19: ...hock disconnect the Agilent Technologies model product from mains before cleaning Use a dry cloth or one slightly dampened with water to clean the external case parts Do not attempt to clean internally Declaration of Conformity For a copy of the manufacturer s Declaration of Conformity for this apparatus contact your local Agilent Technologies office or sales representative Refer to Contacting Agi...

Page 20: ...orme NMB du Canada Compliance with German Noise Requirements This is to declare that this instrument is in conformance with the German Regulation on Noise Declaration for Machines Laermangabe nach der Maschinenlaermrerordnung 3 GSGV Deutschland Acoustic Noise Emission Geraeuschemission LpA 70 dB Lpa 70 dB Operator Position am Arbeitsplatz Normal Operation normaler Betrieb per ISO 7779 nach DIN 456...

Page 21: ... same type and rating United States F 0 5A 250V Part Number 2110 0202 Europe F 3 15A 250V Part Number 2110 0655 The use of other fuses or material is prohibited WARNING This is a Safety Class I product provided with a protective earthing ground incorporated in the power cord The mains plug shall be inserted only into a socket outlet provided with a protective earth contact Any interruption of the ...

Page 22: ...e CAUTION Always use the three prong ac power cord supplied with this instrument Failure to ensure adequate earth grounding by not using this cord can cause instrument damage CAUTION This product is designed for use in Installation Category II and Pollution Degree 2 per IEC 61010 1 2001 CAUTION This instrument has autoranging line voltage input be sure the supply voltage is within the specified ra...

Page 23: ...symbol indicates separate collection for electrical and electronic equipment mandated under EU law as of August 13 2005 All electric and electronic equipment are required to be separated from normal waste for disposal Reference WEEE Directive 2002 96 EC This text indicates that the instrument is an Industrial Scientific and Medical Group 1 Class A product CISPR 11 Clause 4 This symbol indicates th...

Page 24: ... Internet at http www agilent com find assist You can also purchase accessories or documentation items on the Internet at http www agilent com find If you do not have access to the Internet contact your field engineer NOTE In any correspondence or telephone conversation refer to the product by its model number and full serial number With this information the Agilent representative can determine wh...