Chapter 5
205
Language Reference
CALibration Subsystem
Align the Narrow Crystal Prefilter
:CALibration:PFILter:XTALNarrow
:CALibration:PFILter:XTALNarrow?
Align the narrow crystal prefilter. (2.5 kHz to 20 kHz)
Remarks:
A valid service password needs to be entered prior to
sending the command.
Front Panel
Access:
Enter service password and press
System, Diagnostics
Adjust the Level of the 321.4 MHz Alignment Signal
:CALibration:REF321
:CALibration:REF321?
Calculate the curve coefficients for setting the level of the 321.4 MHz
alignment signal.
Remarks:
A valid service password needs to be entered prior to
sending the command.
Front Panel
Access:
System, Diagnostics
50 MHz Reference Alignment Signal
Process
Process Step Description
Command
Both
Attach a 50 MHz signal to the RF input.
Automatic
Does the entire procedure
CAL:REF50[:DOIT]
Interactive
Enter the interactive mode
CAL:REF50:ENTer
Interactive
Tell the instrument what the external
signal’s power is. (approx. -25 dBm)
CAL:REF50:AMPL
Interactive
Proceed with the adjustment phase.
CAL:REF50:ANOW
Interactive
Exit from the interactive mode.
CAL:REF50:EXIT
Query
Return the last alignment value of the
absolute level of the 50 MHz cal signal.
CAL:REF50:LAST:ABSLevel?
Query
Return the last alignment value of the
ALC DAC.
CAL:REF50:LAST:ALCDac?
Summary of Contents for E4406A VSA Series
Page 4: ...4 ...
Page 59: ...59 2 Programming Fundamentals ...
Page 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Page 125: ...125 3 Programming Examples ...
Page 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Page 165: ...165 4 Programming Command Cross References ...
Page 379: ...379 6 Error Messages ...
Page 412: ...412 Chapter6 Error Messages Error Message Descriptions ...