244
Chapter 5
Language Reference
MEASure Group of Commands
50 MHz Amplitude Reference Measurement
This aligns the internal 50 MHz reference signal to an external
reference signal that you supply. You must be in the Service mode to use
these commands. Use INSTrument:SELect to set the mode.
The general functionality of CONFigure, FETCh, MEASure, and READ
are described at the beginning of this section. See the
SENSe:AREFerence commands for more measurement related
commands.
:CONFigure:AREFerence
:FETCh:AREFerence[n]?
:READ:AREFerence[n]?
:MEASure:AREFerence[n]?
Remarks:
For auto adjustment of the internal 50 MHz amplitude
reference, use
CALibration:AMPLitude:REFerence:AADJust
command after this measurement has been selected.t
Front Panel
Access:
Measure, 50 MHz Amptd
After the measurement is selected, press
Restore Meas
Defaults
to restore factory defaults.
Measurement Results Available
n
Results Returned
not specified or n=1
Returns 7 scalar results:
1. RF input average amplitude
2. 50 MHz reference oscillator average amplitude
3. Average amplitude error
4. State (for factory use only)
5. Level (for factory use only)
6. Monitored level (for factory use only)
7. Connector status (for factory use only)
2
RF input amplitude trace data.
3
50 MHz oscillator amplitude trace data
4
Amplitude error strip chart trace data
Summary of Contents for E4406A VSA Series
Page 4: ...4 ...
Page 59: ...59 2 Programming Fundamentals ...
Page 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Page 125: ...125 3 Programming Examples ...
Page 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Page 165: ...165 4 Programming Command Cross References ...
Page 379: ...379 6 Error Messages ...
Page 412: ...412 Chapter6 Error Messages Error Message Descriptions ...