Chapter 5
Language Reference
MEASure Group of Commands
not specified or
n=1
Returns the following comma-separated scalar results:
1. FFT peak is the FFT peak amplitude.
2. FFT frequency is the FFT frequency of the peak amplitude.
3. FFT points is the Number of points in the FFT spectrum.
4. First FFT frequency is the frequency of the first FFT point of the
spectrum.
5. FFT spacing is the frequency spacing between the FFT points of the
spectrum.
6. Time domain points is the number of points in the time domain trace
used for the FFT. The number of points doubles if the data is complex
instead of real. See the time domain scaler description below.
7. First time point is the time of the first time domain point, where time
zero is the trigger event.
8. Time spacing is the time spacing between the time domain points. The
time spacing value doubles if the data is complex instead of real. See the
time domain scaler description below.
9. Time domain returns a 1 if time domain is complex (I/Q) and complex
data will be returned. It returns a 0 if the data is real. (raw ADC
samples) When this value is 1 rather than 0 (complex vs. real data), the
time domain points and the time spacing scalers both increase by a
factor of two.
10. Scan time is the total scan time of the time domain trace used for the
FFT. The total scan time = (time spacing)
Χ
(time domain points
−
1)
11. Current average count is the current number of data measurements
that have already been combined, in the averaging calculation.
2,
Service
mode
only
Returns the trace data of the log-magnitude versus time. (That is, the RF
envelope.)
3
Returns the I and Q trace data. It is represented by I and Q pairs (in volts)
versus time.
4
Returns spectrum trace data. That is, the trace of log-magnitude versus
frequency. (The trace is computed using a FFT.)
5,
Service
mode
only
Returns the averaged trace data of log-magnitude versus time. (That is, the
RF envelope.)
6
Not used.
7
Returns the averaged spectrum trace data. That is, the trace of the
averaged log-magnitude versus frequency.
8
Not used.
9,
Service
mode
only
Returns a trace containing the shape of the FFT window.
n
Results Returned
Summary of Contents for E4406A VSA Series
Page 4: ...4 ...
Page 59: ...59 2 Programming Fundamentals ...
Page 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Page 125: ...125 3 Programming Examples ...
Page 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Page 165: ...165 4 Programming Command Cross References ...
Page 379: ...379 6 Error Messages ...
Page 412: ...412 Chapter6 Error Messages Error Message Descriptions ...