Chapter 5
289
Language Reference
SENSe Subsystem
Adjacent Channel Power—Power Reference
[:SENSe]:ACP:TYPE PSDRef|TPRef
[:SENSe]:ACP:TYPE?
Selects the measurement type. This allows you to make absolute and
relative power measurements of either total power, or the power
normalized to the measurement bandwidth.
Power Spectral Density Reference (PSDRef) - the power spectral
density is used as the power reference
Total Power Reference (TPRef) - the total power is used as the power
reference
Factory Preset
and *RST:
Total power reference (TPRef)
Remarks:
You must be in the Basic, cdmaOne, cdma2000,
W-CDMA (3GPP), W-CDMA (Trial & Arib), NADC, or
PDC mode to use this command. Use
INSTrument:SELect to set the mode.