312
Chapter 5
Language Reference
SENSe Subsystem
Power vs. Time (Burst Power) Measurement
Commands for querying the power versus time measurement results
and for setting to the default values are found in the
. The equivalent front panel keys for the
parameters described in the following commands, are found under the
Meas Setup
key, after the
Power vs Time
measurement has been selected
from the
MEASURE
key menu.
Power vs. Time—Number of Bursts Averaged
[:SENSe]:PVTime:AVERage:COUNt <integer>
[:SENSe]:PVTime:AVERage:COUNt?
Set the number of bursts that will be averaged. After the specified
number of bursts (average counts), the averaging mode (terminal
control) setting determines the averaging action.
Factory Preset
and *RST:
10
Range:
1 to 10,000
Remarks:
You must be in the EDGE(w/GSM), GSM or Service
mode to use this command. Use INSTrument:SELect to
set the mode.
Power vs. Time—Averaging State
[:SENSe]:PVTime:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:PVTime:AVERage[:STATe]?
Turn averaging on or off.
Factory Preset
and *RST:
Off
Remarks:
You must be in the EDGE(w/GSM), GSM or Service
mode to use this command. Use INSTrument:SELect to
set the mode.
Summary of Contents for E4406A VSA Series
Page 4: ...4 ...
Page 59: ...59 2 Programming Fundamentals ...
Page 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Page 125: ...125 3 Programming Examples ...
Page 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Page 165: ...165 4 Programming Command Cross References ...
Page 379: ...379 6 Error Messages ...
Page 412: ...412 Chapter6 Error Messages Error Message Descriptions ...