406
Chapter 6
Error Messages
Error Message Descriptions
GSM - Specific Error Messages
[100 to 199]
An error number in the range [100 to 199] indicates the instrument has
detected an error relating to the GSM personality.
GSM - Specific Error Message Descriptions
(Number)
Description
(100)
Not enough data to fit into GSM mask
An attempt to position a GSM trace into the mask,
when not enough data was present. Try using the
Restart
key to clear the problem. This can be caused by
a bad GSM burst, or the RF Sync Delay set too far.
(101)
GSM burst out of limits
The GSM signal did not fit into the mask in the Power
vs. Time measurement.
(102)
Insufficient pre-Trig for demod - decrease
Trig Delay
(103)
Incorrect RBW for demod - change RBW
(104)
Invalid GSM burst timing
A GSM-like burst was acquired, but it’s timing is not
valid. Ensure the correct
Burst Type
has been selected.
(105)
Valid GSM burst not found
In a GSM measurement, data was acquired but a GSM
burst was not found.
(106)
Cannot synchronize frame trigger
Cannot synchronize the frame trigger to the even
second clock.
(107)
Dynamic range not optimum - set AUTO RF input
(108)
Cannot synchronize to RF amplitude (burst
error)
(109)
GSM RF sync delay is out of range
Change RF Sync Delay.
Summary of Contents for E4406A VSA Series
Page 4: ...4 ...
Page 59: ...59 2 Programming Fundamentals ...
Page 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Page 125: ...125 3 Programming Examples ...
Page 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Page 165: ...165 4 Programming Command Cross References ...
Page 379: ...379 6 Error Messages ...
Page 412: ...412 Chapter6 Error Messages Error Message Descriptions ...