CALibration Subsystem
Internal 50 MHz Amplitude Reference Alignment Control
:CALibration:REF50[:DOIT]
:CALibration:REF50[:DOIT]?
Does automatic alignment of the internal 50 MHz amplitude reference
oscillator. You do this by setting an external source to
−
25.00 dBm and
using a power meter to measure the exact value. Then use
CAL:REF50:AMPL to input the source amplitude, measured on the
power meter. Finally, connect the source to the instrument RF INPUT
port and run the adjustment.
Remarks:
You must be in the Service mode to use this command.
Use INSTrument:SELect.
A valid service password needs to be entered prior to
sending this command
Front Panel
Access:
System
,
Alignments, Align subsystem
,
Align 50 MHz
Reference
Enter Interactive Mode for Internal 50 MHz Amplitude
Reference Alignment
:CALibration:REF50:ENTer
Turns on the interactive mode for alignment of the internal 50 MHz
amplitude reference signal. Use CAL:REF50:ANOW to do the
alignment and CAL:REF50:EXIT to exit the interactive mode.
Remarks:
You must be in the Service mode to use this command.
Use INSTrument:SELect.
A valid service password needs to be entered prior to
sending this command.
Front Panel
Access:
System
,
Alignments, Align subsystem
,
Align 50 MHz
Reference
Summary of Contents for E4406A VSA Series
Page 4: ...4 ...
Page 59: ...59 2 Programming Fundamentals ...
Page 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Page 125: ...125 3 Programming Examples ...
Page 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Page 165: ...165 4 Programming Command Cross References ...
Page 379: ...379 6 Error Messages ...
Page 412: ...412 Chapter6 Error Messages Error Message Descriptions ...