248
Chapter 5
Language Reference
MEASure Group of Commands
Power vs. Time Measurement
This measures the average power during the “useful part” of the burst
comparing the power ramp to required timing mask. You must be in
EDGE, GSM or Service mode to use these commands. Use
INSTrument:SELect to set the mode.
The general functionality of CONFigure, FETCh, MEASure, and READ
are described at the beginning of this section. See the SENSe:PVTime
commands for more measurement related commands.
:CONFigure:PVTime
:FETCh:PVTime[n]?
:READ:PVTime[n]?
:MEASure:PVTime[n]?
Front Panel
Access:
Measure, Power vs Time
After the measurement is selected, press
Restore Meas
Defaults
to restore factory defaults.
Measurement Results Available
n
Results Returned
0
Returns unprocessed I/Q trace data, as a series of comma-separated trace
points, in volts. The I values are listed first in each pair, using the 0 through
even-indexed values. The Q values are the odd-indexed values.
Summary of Contents for E4406A VSA Series
Page 4: ...4 ...
Page 59: ...59 2 Programming Fundamentals ...
Page 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Page 125: ...125 3 Programming Examples ...
Page 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Page 165: ...165 4 Programming Command Cross References ...
Page 379: ...379 6 Error Messages ...