270
Chapter 5
Language Reference
SENSe Subsystem
Remarks:
You must be in Basic, cdmaOne, cdma2000, W-CDMA
(3GPP), W-CDMA (Trial & Arib), or iDEN mode to use
this command. Use INSTrument:SELect to set the
mode.
Adjacent Channel Power—FFT Segments
[:SENSe]:ACP:OFFSet:LIST:FFTSegment
<integer>,<integer>,<integer>,<integer>,<integer>
[:SENSe]:ACP:OFFSet:LIST:FFTSegment?
Selects the number of FFT segments used in making the measurement.
In automatic mode the measurement optimizes the number of FFT
segments required for the shortest measurement time. The minimum
number of segments required to make a measurement is set by your
desired measurement bandwidth. Selecting more than the minimum
number of segments will give you more dynamic range for making the
measurement, but the measurement will take longer to execute.
Factory Preset
and *RST:
Range:
1 to 12
Remarks:
You must be in Basic, cdmaOne mode to use this
command. Use INSTrument:SELect to set the mode.
History:
Revision A.03.00 or later, in cdmaOne revision A.04.00
Mode
Offset A
Offset B
Offset C
Offset D
Offset E
Basic &
cdmaOne
1
1
1
1
1
Summary of Contents for E4406A VSA Series
Page 4: ...4 ...
Page 59: ...59 2 Programming Fundamentals ...
Page 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Page 125: ...125 3 Programming Examples ...
Page 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Page 165: ...165 4 Programming Command Cross References ...
Page 379: ...379 6 Error Messages ...
Page 412: ...412 Chapter6 Error Messages Error Message Descriptions ...