Chapter 5
285
Language Reference
SENSe Subsystem
Remarks:
You must be in Basic, cdmaOne, cdma2000, W-CDMA
(3GPP), W-CDMA (Trial & Arib), or iDEN mode to use
this command. Use INSTrument:SELect to set the
mode.
Adjacent Channel Power—Number of Measured Points
[:SENSe]:ACP:POINts <integer>
[:SENSe]:ACP:POINts?
Selects the number of data points used to measure the reference
(carrier) channel. The automatic mode chooses the optimum number of
points for the fastest measurement time with acceptable repeatability.
The minimum number of points that could be used is determined by the
sweep time and the sampling rate.
You can increase the length of the measured time record (capture more
of the burst) by increasing the number of points, but the measurement
will take longer. Use
[:SENSe]:ACP:OFFSet:LIST:POINts
to set the
number of points used for measuring the offset channels.
Factory Preset
and *RST:
1024
Remarks:
The fastest measurement times are obtained when the
number of points measured is 2
n
.
You must be in Basic, cdmaOne mode to use this
command. Use INSTrument:SELect to set the mode.
Range:
64 to 65536
W-CDMA
(3GPP)
REL
REL
REL
REL
REL
W-CDMA (Trial
& Arib)
REL
REL
REL
REL
REL
Mode
Variant
Offset A
Offset B
Offset C
Offset D
Offset E
Summary of Contents for E4406A VSA Series
Page 4: ...4 ...
Page 59: ...59 2 Programming Fundamentals ...
Page 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Page 125: ...125 3 Programming Examples ...
Page 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Page 165: ...165 4 Programming Command Cross References ...
Page 379: ...379 6 Error Messages ...
Page 412: ...412 Chapter6 Error Messages Error Message Descriptions ...